Aeroneb
®
Solo System Instruction Manual
15
Nebulization Modes
Intermittent Mode
Warnings
•
To avoid damage to the Aeroneb Solo, do not use a syringe with needle.
•
During use observe for correct functioning of the nebulizer.
•
The maximum capacity of the nebulizer is 6 mL.
For intermittent doses less than or equal to 6 mL:
1. Open the plug on the nebulizer.
2. Use a pre-filled ampoule or syringe to add medication into the filler port of the
nebulizer (Figure 11).
3. Close the plug.
Figure 11.
Filling the nebulizer with a pre-filled ampoule
4. To start a 30 minute nebulization cycle, press and release the blue on/off power
button (Figure 4). The green 30 minute indicator light illuminates to indicate that
the 30 minute nebulization cycle is in progress.
5. To stop the nebulizer at any time, press the on/off power button. The indicator
turns off to indicate that nebulization has stopped.
Note:
Medication can be added to the Aeroneb Solo during nebulization. This does
not interrupt nebulization or ventilation.