8
Aerogen
Aeroneb Pro-X Controller
Figure 4.
Aeroneb Pro-X Controls & indicators
Table 1.
Aeroneb Pro-X Controls & Indicators
Control / Indicator
Function
30 Min. Indicator
•
Green (steadily lit) = 30 minute nebulization cycle on
•
Green (flashing) = Low battery power
•
Nebulizer automatically powers off after 30 minutes have elapsed
Continuous Indicator
•
Green (steadily lit) = Continuous nebulization cycle on
•
Nebulizer does not power off automatically
Fault Indicator
•
Amber (steadily lit) = Aeroneb Solo nebulizer disconnected from
Aeroneb Pro-X controller
•
Amber (flashing) = Aeroneb Pro-X drive voltage error
On/Off Power Button
•
To operate in 30 minute mode press the On/Off button once
•
To operate in continuous mode press and hold the On/Off button for
greater than 3 seconds from off
•
Pressing during nebulization turns off power to the nebulizer
Battery Status
Indicator
•
Green = Battery fully charged.
•
Amber = Battery charging.
•
No light = Battery in operation.
Fault Indicator
30 Minute
Indicator
Continuous
Mode Indicator
Controller
Cable Input
On/Off Power
• 30 Min. - Press & Release
• Continuous Mode - 3 Sec. hold from off
Battery Status
Indicator
9V DC Input