2. The Aeroneb Solo can be placed on the dry side of the humidifier as shown in
Figure 7. The Aeroneb Solo can be used with a nasal interface in this configuration.
Figure 7.
Aeroneb Solo on dry side of humidifier
3. Follow ventilator manufacturer instructions for performing a leak test after
inserting or removing the nebulizer.
Warnings
•
Condensate can collect and occlude ventilator circuits. Always position ventilator
circuits so that fluid condensate drains away from the patient.
•
Always connect a bacteria filter to the expiratory inlet of the ventilator. Otherwise
the function of the expiratory channel may be degraded.
•
Do not use a filter or heat-moisture exchanger (HME) between the nebulizer and
patient airway.