Simple Logger
®
II Series
21
MODEL
L562
ELECTRICAL
Channels
Two
Connection
Current Channel
Voltage Channel
Input
BNC
Two recessed banana jacks
Input Level
0 to 1V (for use with current
probes with a voltage output)
0 to 600V
AC
Accuracy
(50/60Hz)
0 to 10mV unspecified
10 to 50mV:
±(0.5% of R 1mV)
50 to 1000mV:
±(0.5% of R 0.5mV)
0 to 5V unspecified
5 to 50V:
±(0.5% of R 1V)
50 to 600V:
±(0.5% of R 0.5V)
Resolution
0.1mV
0.1V
Maximum Input Voltage***
5Vrms or
±
7.07V
PEAK
1.2 x 600V
Input Impedance
800k
W
40M
W
Sample Rate
64 samples/cycle
Storage Rate
Programmable from 125ms to 1 per day
Storage Modes
Start/Stop, FIFO and Extended Recording Mode
*
(XRM
™
), Alarm
Recording Length
15 minutes to 8 weeks, programmable using DataView
®
Memory
240,000 measurements (512KB). Recorded data is stored in non-volatile
memory and retained even if battery is low or removed.
Communication
USB 2.0 optically isolated
Power Source**
2x1.5V AA-cell alkaline batteries
Battery Life
100 hours to >45 days (dependent on storage rate/recording length)
MECHANICAL
Dimensions
5.38 x 2.75 x 1.28" (136 x 70 x 32mm)
Weight (with battery)
6.4 oz (180g)
Case
UL94-V0
Vibration
IEC 68-2-6 (1.5mm, 10 to 55Hz)
Shock
IEC 68-2-27 (30G)
Drop
IEC 68-2-32 (1m)
ENVIRONMENTAL
Operating Temperature
14° to 122°F (-10° to 50°C)
Storage Temperature
-4° to 140°F (-20° to 60°C)
Relative Humidity
Up to 85% at 95°F (35°C), Non-condensing
Altitude
6562’ (2000m)
SAFETY & ELECTRO-MAGNETIC COMPATIBILITY
Safety Rating
EN61010-1; 600V CAT III, 300V CAT IV; Pollution Degree 2
Protection Degree
IP40
Electro-Magnetic
Compatibility
EN 61326-1; 07/1997 (+A1 10/1998, +A2 09/2001, +A3 05/2004)
CE Approved
Yes