30
Simple Logger
®
II Series
4.4
Downloading Recorded Data
Recorded measurements stored in the instrument are transferred to a
computer via the DataView Simple Logger II Control Panel, as described
in the Help system that comes with the program.
4.5
Clearing Alarm Indication
Clearing alarms can be performed in the
STANDBY
or
RECORD
mode.
1. Press and hold the
PRESS
button. When the
ALARM
(YELLOW) LED
lights up, release the button. The YELLOW LED will blink at a fast rate
for a period of five seconds.
2. Press the
PRESS
button for another 0.5 second to complete the oper
-
ation.
This does not clear any stored alarms, only indications. Stored alarms can
only be cleared when memory is erased (see
§
4.6).
4.6
Erasing Data from Memory
Erasing data from the instrument’s memory can only be performed while
in the STANDBY mode.
There are two ways to erase the memory:
Erasing the Memory using the PRESS Button:
1. Press and hold the
PRESS
button. When the
ERASE
(RED) LED
lights up, release the button.
This will arm the instrument for an erase operation (when not in record
mode). While armed to erase memory, the RED LED will blink at a fast
rate for a period of five seconds.
2. Press the
PRESS
button for another 0.5 second to start the erase
operation. During the erase operation the LED will blink once every
second. Erasure takes about 20 seconds to complete.
www.
GlobalTestSupply
.com
Find Quality Products Online at: