5.2
R3760 Performance Test
5-8
5.2.
R3760 Performance Test
This chapter describes testing procedures designed to maintain this analyzer's efficiency.
For items not covered in this chapter, contact Advantest.
5.2.1. Preparing for a Performance Test
1.
Warm up
Warm up the R3760 for at least 30 minutes before executing the performance test.
2.
Preparing measurement instruments
The following measurement instruments are required to perform the test items as shown in
Use the sample software from the CD-ROM manual (FOC-8440274) for performance verification.
Table 5-2 Required Measurement Instruments for Performance Test
Test items
Measurement instrument
Remarks
1. Frequency accuracy
and range
•
Counter
Frequency:
100 MHz to 6 GHz
Display:
7 digits or more
Accuracy:
0.1ppm or less
•
SMA-SMA cable
2. Output/Input level and
flatness
•
Power meter
Frequency:
100 MHz to 6 GHz
Power range: -20 dBm to +10 dBm
•
Power sensor
Frequency:
100 MHz to 6 GHz
Power range: -43 dBm to +21 dBm
3. Output level
linearity
•
Power meter
Frequency:
100 MHz to 6 GHz
Power range: -20 dBm to +10 dBm
•
Power sensor
Frequency:
100 MHz to 6 GHz
Power range: -20 dBm to +10 dBm
4. Dynamic Level Accu-
racy
•
Step Attenuator
Variable Range: 0 dB - 90 dB
Accuracy:
Within 0.02 dB
•
RF Cable
(SMA(m)/SMA(m) 50
Ω
)
×
2
•
Transformer Connectors
(N(m)/SMA (f))
×
2
•
2 dB Fixed Attenuator
×
1
•
3 dB Fixed Attenuator
×
1
(SMA(f)/SMA(m))