A.3.1
Descriptions of Window Components
A-26
µ
W:
When selected, the level is calculated by
µ
W.
By default, it is calculated by dBm.
Setting:
Calculates the level and sets conditions based on measurement con-
ditions.
Meas:
Performs DLD measurement.
Save Data:
Saves the measurement results to a file in the CSV file format.
NOTE: DLD measurement: An abbreviation for Drive Level Dependency.
A.3.1.7
Display Scale Setting Area
The display format for measurement data can be set to ON or OFF, and scale can be set.
Available display formats are only the logarithmic amplitude, phase, group delay, standing wave ratio, real
part, and imaginary part.
Figure A-31 LOGMAG / PHASE / DELAY / SWR / REAL / IMAG
Each titled field sets the following information:
LOGMAG / DELAY / SWR / REAL / IMAG:
When LOGMAG is displayed, the display is the logarithmic ampli-
tude scale.
When DELAY is displayed, the display is the group delay scale.
When SWR is displayed, the display is the standing wave ratio
scale.
When REAL is displayed, the display is the real number scale.
When IMAG is displayed, the display is the imaginary number
scale.
PHASE:
If this check box is selected, the phase waveform display is enabled.
The display format is set to
±
180
°
C phase shift display.
REF:
Specify a reference level for the scale.
/DIV:
Specify a division of the vertical axis scale of the Cartesian coordi-
nate display, by a numeric value.
POS:
Sets a reference position of the vertical axis. Specify a value from 0
to 100.