A.3.1
Descriptions of Window Components
A-29
A.3.1.10
CAL
Figure A-34 CAL
Various settings associated with calibration can be made and calibration can be executed.
Each titled field sets the following information:
CAL:
This check box is selected during calibration.
If calibration data exists, selecting this check box sets calibration
execution to ON and deselecting this check box sets calibration exe-
cution to OFF.
If calibration data does not exist, this check box is automatically
selected when calibration is executed.
Norm:
Clicking this button calibrates frequency characteristics of ampli-
tude and phase.
This calibration method is referred to as normalization.
clear:
Clicking this button clears calibration data.
The previous calibration data must be cleared when calibration is
executed again.
However, if normalization is executed, it can be made without clear-
ing the previous calibration data.
Method:
Select a calibration method from the drop down menu.
Selectable calibration methods are as follows:
Norm:
Selects normalization.
Trans:
Selects transmission full calibration.
1Port:
Selects 1-port full calibration.
sv:
Clicking this button saves calibration data.
The saved data is cleared when the sample program is terminated.
rc:
Clicking this button calls calibration data.