A.3.1
Descriptions of Window Components
A-18
A.3.1.4
Connection Box
Figure A-21 Connection Box
Each titled field sets the following information:
BD:
Set an R3755x/R3760 board number to be connected, by a value
from 0 to 7.
IP:
When the sample program operates on the host PC, entering a value
is unnecessary.
Directly enter the IP address of the host PC where "NACserver" is
running, or enter the PC name XXXXXXXX of the host PC in the
"XXXXXXXX" format.
Example: When the host PC name is R3755A:
"R3755A"
Connect:
Clicking this button connects the R3755x to the host PC whose IP
address (or PC name) is specified above.
A.3.1.5
Trigger
Figure A-22 Trigger
Each titled field sets the following information:
Cont:
Clicking this button executes sweeps successively.
Hold/Abort:
Clicking this button immediately terminates the sweep.
Single:
Clicking this button executes a single sweep.
If this button is clicked during a sweep, the measurement being
swept is stopped and the sweep is started again.
Part:
If this check box is selected, a sweep is partially executed.
A sweep is executed in a range according to the sweep start and stop
points entered in the text boxes below the check box.
Note that the points specified in those text boxes must be values
from 0 to 1600. In addition, the values must be one or more less than
the total point counts entered in the Point field for individual seg-
ments of the "Stimulus" field.
Example When 201 points are set for each of Stimulus 1 and
Stimulus 2:
Maximum specifiable value = 201 + 201 - 1 = 401
NOTE:
Partial sweeping is enabled only in the Single sweep mode.