Operation
45
USB-1900 Series
USB-1900 Series starts to acquire data. The total acquired data
length = NumChan_counter * PSC_counter.
Figure 4-8: Delayed Trigger
4.4.3
Post-Trigger or Delayed-Trigger Acquisition with
Retriggering
Post-trigger or Delayed-Trigger acquisition with re-trigger function
is indicated where you want to collect data after several trigger
events. The number of scans after each trigger is specified in
PSC_counter, and Retrig_no can be programmed to specify the
re-trigger numbers. An example is shown, in which two scans of
data are acquired after the first trigger signal, after which the
USB-1900 Series waits for the next re-trigger signal (re-trigger sig-
WARNING:
When the Delay_counter clock source is set to timebase, the
maximum delay time = 2
32
/80M s = 18.626ms
(NumChan _Counter=4, PSC_Counter=3)
Acquisition_in_progress
Scan_start
AD_conversion
Acquired & Stored Data
(3 scans)
Operation start
Trigger
Delay until
Delay_Counter
reaches 0
Summary of Contents for USB-1901
Page 6: ...vi This page intentionally left blank...
Page 10: ...x This page intentionally left blank...
Page 14: ...xiv This page intentionally left blank...
Page 24: ...10 Introduction This page intentionally left blank...
Page 26: ...12 Hardware Information Figure 2 2 USB 1902 Module Side View...
Page 27: ...Hardware Information 13 USB 1900 Series Figure 2 3 USB 1902 Module Front View 41 3 114...
Page 29: ...Hardware Information 15 USB 1900 Series Figure 2 6 Module in Stand Front View 114 3...
Page 30: ...16 Hardware Information Figure 2 7 Module Stand Top View 20 4 20 4 B 26...
Page 33: ...Hardware Information 19 USB 1900 Series Figure 2 12 Module Rail Mounted...
Page 44: ...30 Hardware Information This page intentionally left blank...
Page 48: ...34 Installing the USB 1900 Series Module This page intentionally left blank...
Page 78: ...64 Operation This page intentionally left blank...