66
Calibration
5.3
Saving Calibration Constants
Factory-calibrated constants are permanently stored in a bank of
the onboard EEPROM and cannot be modified. When the device
is recalibrated, the software stores the new constants in a user-
configurable section of the EEPROM. To restore original factory
calibration settings, the software can copy the factory-calibrated
constants to the user-configurable section of the EEPROM. When
auto-calibration is complete, the new calibration constants can be
saved to the user-configurable banks in the EEPROM.
NOTE:
NOTE:
The USB-1903, requiring an external precision current source
to calibrate the current-to-voltage conversion resistor and the
differential buffer in its front-end circuit, does not support auto-
calibration. Please return the module for calibration service if
necessary.
NOTE:
NOTE:
X
The USB-1900 Series should be warmed up for
at least 15 minutes before initiating auto-calibra-
tion
X
Remove the cable before auto-calibrating the
card since the DA outputs are changed during
the process
Summary of Contents for USB-1901
Page 6: ...vi This page intentionally left blank...
Page 10: ...x This page intentionally left blank...
Page 14: ...xiv This page intentionally left blank...
Page 24: ...10 Introduction This page intentionally left blank...
Page 26: ...12 Hardware Information Figure 2 2 USB 1902 Module Side View...
Page 27: ...Hardware Information 13 USB 1900 Series Figure 2 3 USB 1902 Module Front View 41 3 114...
Page 29: ...Hardware Information 15 USB 1900 Series Figure 2 6 Module in Stand Front View 114 3...
Page 30: ...16 Hardware Information Figure 2 7 Module Stand Top View 20 4 20 4 B 26...
Page 33: ...Hardware Information 19 USB 1900 Series Figure 2 12 Module Rail Mounted...
Page 44: ...30 Hardware Information This page intentionally left blank...
Page 48: ...34 Installing the USB 1900 Series Module This page intentionally left blank...
Page 78: ...64 Operation This page intentionally left blank...