Software Utility
8.3 Calibration
This function will guide you on how to calibrate the PCI-9812/10.
The calibration program serves as a useful test for the PCI-
9812/10's A/D,D/A and DIO functions and can aid in troubleshooting
if problems arise.
Note :
For an environment with frequent large changes in temperature
and vibration, a recalibration interval, 3 months, is
recommended. For laboratory conditions, 6 months to 1 year is
acceptable
When you choose the calibration function from the main menu list,
The calibration item menu is displayed on the screen. After you
select one of the calibration item from the menu, a calibration
window will appeor. The upper window shows the detailed
procedures which you have to follow when proceeding the
calibration. The instructions will tell you how to calibrate each item
step by step. The bottom window shows the layout of PCI-9812/10.
In addition, the proper Variable Resister (VR) will blink to indicate
the related VR needs to be adjusted for the current calibration step.
****** PCI-9812 Calibration ******
<1> A/D channel 0 adjusting
<2> A/D channel 1 adjusting
<3> A/D channel 2 adjusting
<4> A/D channel 3 adjusting
<Esc> Quit
Select 1 to 4 or <Esc> to quit calibration.