GB 02923 - Edition 05 - 03/2013
1/5
B 300
Pump connections to an installation
Equipment
installation conditions
The pump can operate in
any position
When the pump is connected according to the manufacturer recommendations
(refer to B300), in case of overpressure resulting from an incident, the pump
enclosure can withstand a maximum static pressure of 1.1 MPa for 1 minute.
Pump installation with inlet housing facing down.
- connect the pump to the customer’s handling device using the threaded
holes located at the rear of the pump (refer to chapter A400 of the operating
manual).
The equipment frame on which the pump is installed must be sufficiently
rigid to absorb the kinetic energy of the rotor in case of pump rotor crash.
For this, take into account :
• the maximum loads to calculate the equipment attachment devices,
• the flange dimensions,
• the quality and the number of bolts.
Do not use bellows on the pump inlet flange and the chamber Reducing
adapter must be correctly dimensioned.
In.
In.
In.
In.
adixen Vacuum Products - Operating Instructions - ATP 2300 M