61
•
Some input file fields could be empty. However, last name, first name and date of birth fields are
mandatory.
Example:
Smith,John,2016-02-24,,,,0
is a valid input file.
Output file specifications
The output file presents information in a CSV file. Comma separators are used. There are no headers in the
first line. Information of each measurement/screening is presented side by side in a single line in this
sequence.
For refraction, always positive cylinder notation is used.
Data in the output file are presented in the following order:
•
Date and time stamp, formatted as YYYY-MM-DDThh:mm:ss where T is a separator
•
Same last name as provided in input file
•
Same first name as provided in input file
•
Same date of birth as provided in input file
•
Same patient ID as provided in input file
•
Same location as provided in input file
•
Same contact as provided in input file
•
Spherical measurement value of the right eye in diopters [dpt]
•
Cylindrical measurement value of the right eye in diopters [dpt]
•
Axis measurement value of the right eye in degrees [deg]
•
Pupil diameter measurement value of the right eye in millimeters [mm]
•
Spherical measurement value of the left eye in diopters [dpt]
•
Cylindrical measurement value of the left eye in diopters [dpt]
•
Axis measurement value of the left eye in degrees [deg]
•
Pupil diameter measurement value of the left eye in millimeters [mm]
•
Measurement value of gaze x value for right eye in degrees [deg]
•
Measurement value of gaze y value for right eye in degrees [deg]
•
Measurement value of gaze x value for left eye in degrees [deg]
•
Measurement value of gaze y value for left eye in degrees [deg]
•
Distance between left and right pupil in millimeters [mm]
•
2WIN exam number [int]
•
2WIN software version [string]
•
Exam ending message in case of inconclusive result [string]
•
Exam ending message code [int] (if 0 everything is ok).