19
To operate correctly the instrument, it
’
s important to carry out the following main steps:
1.
Device setup
2.
Preparation of the patient
3.
Measurement
9.1 Device setup
To charge the system:
1.
Connect the provided power supply cable to the
INPUT 18V
–
25A,
depicted in figure.
2.
Connect the other end of the power supply to the line.
3.
In order to monitor the battery charge level, press the battery charge
button.
Configuration steps
Refer to Chapter 11 of this Operator Manual for 2WIN-S configuration using the Wi-Fi tablet application
(deprecated legacy feature).
The 2WIN-S configuration using the Bluetooth tablet application (KALEIDOS App) is briefly described in the
following.
At first use, it is necessary to pair the 2WIN-S to the tablet, following these instructions:
•
Install the KALEIDOS App on your tablet
•
Check that Bluetooth is enabled on the tablet
•
Switch the 2WIN-S on
•
Run KALEIDOS App on tablet
•
Tap on the Bluetooth icon on the KALEIDOS App to start available Bluetooth devices search. A window
showing all the available 2WIN-S will appear.
•
Connect to the desired 2WIN-S. When the pairing is complete a confirmation message will appear.
Refer to
Chapter 12
for further information.
In the 2WIN-S
Plus
configuration, the provided tablet is already paired with the device.
For the 2WIN-S configuration and how to assembly the instrument, refer to the
2WIN-S Quick Start
Guide