49
Pressing the VALIDATE button you can check if the inserted data are correct.
Pressing the SAVE button you will confirm the Shared Folder configuration.
5.
Setup your EMR system so that it generates a CSV input file in the Shared Folder, containing the
selected patient
’
s data. Refer to
Appendix B
for the CSV input file specifications.
6.
Setup your EMR system so that it accepts data from the CSV output file created by KALEIDOS App
after the measurement. Refer to
Appendix B
for the CSV output file specifications.
Once the EMR connection setup is completed, follow these steps to start a measurement:
1.
Select a patient in your EMR system. The CSV input file is instantly created in the Shared Folder.
2.
Start a new 2WIN-S (Refraction) measurement from the KALEIDOS App. The app imports the patient
data from the CSV input file and displays it. After the import is completed, the app will delete the
CSV input file from the Shared Folder.
KALEIDOS App is checking the Shared Folder approximately every 2 sec. for a new input file.
3.
If the patient data displayed in KALEIDOS App are correct, start the measurement.
4.
Save the exam. The CSV output file containing the exam results is created in the Shared Folder and it
is retrieved by the EMR system. Then the EMR system deletes the CSV output file.
5.
Review the patient
’
s exam results on the EMR system.
A CSV input file triggers an output file. If the EMR system does not provide an input file to begin with,
your device will not write output files: the Shared Folder will remain empty.
Once a CSV input file is provided, the entry of patient data in the KALEIDOS App is disabled.
Once a CSV input file is provided, the exam will start when the home page of the app is displayed.
You can manage multiple 2WIN-S on the same EMR system. Different 2WIN-S devices work on
different Shared Folders, so you must create a dedicated Shared Folder for each 2WIN-S.