Acromag microBlox uBSP-P-1 User Manual Download Page 17

microBlox® Signal Splitter 

Model uBSP-P-1 

 

 

Acromag, Inc. Tel: 248-295-0880  

            - 17 -                                   

http://www.acromag.com  

- 17 - 

https://www.acromag.com 

 

Quick Overview 

 Android Reconfiguration...continued

 

 

Output Calibration (Each of Two Outputs) 

Scroll down the Calibration page to access the Output Calibration controls:  
output channel selector, adjustment controls, and the 

[CALIBRATE 

OUTPUT ZERO]

 and 

[CALIBRATE OUTPUT FULL-SCALE]

 buttons. 

 
First select the Output channel to calibrate, and its output range will be 
displayed along with some instructions on how to proceed. 
 
For Output Zero calibration, use the output adjustment slider and the 
coarse and fine adjustment controls to precisely set your output zero level 
while precisely monitoring your output signal.  Be sure to use a meter with 
an accuracy at least 4x greater than the signal you are measuring for best 
results.  Note that the output adjustment controls temporarily remove 
control of the output from the input to accomplish calibration (control of 
the output level returns to the input signal after 30 seconds). 
 
Once your output level is precisely set to its zero point (4.000mA for this 
splitter), tap the 

[CALIBRATE OUTPUT ZERO]

 button one time to set the 

output DAC level (its corresponding digital count) to correspond to the 
zero (0%) of your output range. 
 

 

For Full-Scale calibration, use the output adjustment slider and the coarse 
and fine adjustment controls to precisely set your output full-scale level 
while precisely monitoring your output signal.  Be sure to use a meter with 
an accuracy at least 4x greater than the signal you are measuring for best 
results.  Note that the output adjustment controls temporarily remove 
control of the output from the input level to accomplish calibration 
(control of the output level returns to the input signal after 30 seconds). 
 
Once your output level is precisely set to its full-scale level (20.000mA for 
this splitter), tap the 

[CALIBRATE OUTPUT FULL-SCALE]

 button one time to 

set the output DAC level (its corresponding digital count) to correspond to 
the full-scale (100%) level of the output range. 
 
Repeat the Output Calibration of zero and full-scale for the second output 
as required by selecting the opposite channel. 
 
If following calibration, your output acts erratic or appears imprecise, you 
may need to repeat input or output calibration, being very careful to take 
accurate measurements and input correct signal levels.  If you are 
measuring voltage across an output load resistance to measure the 
current level (recommended), make sure that you use exact resistance 
when calculating the measured loop current.  When rescaling I/O, make 

sure that you have adequate I/O span, as “too

-

tight” input or output spans 

will have diminished resolution and magnify error. 

 

 

 

Summary of Contents for microBlox uBSP-P-1

Page 1: ...uB Field Input Modules USER S MANUAL ACROMAG INCORPORATED Tel 248 295 0880 30765 South Wixom Road Fax 248 624 9234 Wixom MI 48393 2417 U S A email sales acromag com Copyright 2018 Acromag Inc Printed...

Page 2: ...ections Each Splitter Output Supports Voltage or Current 11 AUX Voltage Output Connections 12 CONFIGURATION SOFTWARE 12 CONFIGURATION SOFTWARE 13 Quick Overview Android Reconfiguration 13 Quick Overvi...

Page 3: ...or human life is involved It is important that you perform satisfactory overall system design and it is agreed between you and Acromag that this is your responsibility The information of this manual...

Page 4: ...lox module carrier adds wide input isolated power cold junction compensation for thermocouple modules dual isolated current and or voltage outputs an AUX voltage output plus LED indication for module...

Page 5: ...than 0 1 output accuracy Wide ambient temperature operation from 40 C to 75 C All I O and power ports are transient protected Wide range isolated DC power input from 6 32V LED indicators for module B...

Page 6: ...a Windows based personal computer laptop or a wired USB OTG connection to an Android smartphone tablet This carrier provides isolated host power to the module a host USB port a host voltage output th...

Page 7: ...nd lift it from the rail To attach the carrier to T type DIN rail position the top groove of the DIN Clip along the rail and push the bottom inward to snap it into place on the DIN rail To remove it f...

Page 8: ...te 16 26 AWG 1 31 0 13mm2 solid or stranded wire with a minimum temperature rating of 85o C Input wiring may be shielded or unshielded type twisted pair or shielded twisted pair input wiring is recomm...

Page 9: ...function as a simple single channel I O carrier for Bluetooth wireless technology modules and you may configure a B module on this carrier to drive its single AUX output using Bluetooth wireless techn...

Page 10: ...ual 8501 043 uB34 2 3 Wire uB35 4 Wire RTD Input User s Manual 8501 044 Introduction to Agility Configuration Tool for microBlox 8501 047 uB37 uB47 TC Input User Manual 8501 050 uB04 uB08 uB16 Back Pa...

Page 11: ...ected in their connection diagrams for the specific plug in module model Ground connections noted are recommended for best results and help protect the unit and its isolated circuits by giving each is...

Page 12: ...ers laptops or a USB OTG connection to Android smartphones tablets with the Agility APP The Windows USB software for your model operates like the Agility APP software Below is a brief introduction to...

Page 13: ...he device indicated to the right of Selected Device and move to the main portion of the app shown in the second screen at left Note Android requires user permission to access external hardware If the...

Page 14: ...nostic exchange and review the results returned If one or more tests indicate Failed you can also tap the down arrow message below the self test report to access additional information regarding faile...

Page 15: ...akin to a power on reset of the splitter You can tap RESET FACTORY CALIBRATION to get out of trouble if you ever miscalibrate a splitter this only affects splitter calibration You can tap RESTORE FAC...

Page 16: ...ents Input calibration controls first followed by Output calibration controls as you scroll down the page IMPORTANT The splitter has had its input output channels factory calibrated with high precisio...

Page 17: ...our output range For Full Scale calibration use the output adjustment slider and the coarse and fine adjustment controls to precisely set your output full scale level while precisely monitoring your o...

Page 18: ...The input type of the plug in module is shown at the top of the screen actual input value not the scaled input value is polled Select the Indicator pointer to set your desired indication to Digital v...

Page 19: ...nication Set up I O Config Test and Calibration A short description of each of these pages follows This section is used to select a connected transmitter and open close communications with it Device c...

Page 20: ...ich output of this splitter you are calibrating Input First set the Input Range to calibrate in the I O Config Test page and be sure to click the Send I O Config button before attempting calibration O...

Page 21: ...us feature of the uBSP P 1 splitter and its USB software is that it allows all versions of microBlox input modules to be reconfigured using a wired USB connection both B and fixed range uB models exce...

Page 22: ...range zero full scale endpoints I O scaling in a channel can be done in reverse to produce a reverse acting output signal Refer to the block diagram to gain a better understanding of how this splitter...

Page 23: ...fect on standard microBlox modules Unit Fails to Start up or Initialize Check power voltage level and capacity making sure that it is at least 6V and of sufficient capacity select a current capacity 2...

Page 24: ...Check CJC Setting Refer to your module specifications and determine the required CJC status for your model Enable or disable CJC as required for your module by setting the channel CJC DIP switch acco...

Page 25: ...his isolator and cable are also included in TT SIP see above USB A B Cable USB A B Cable Order 4001 112 This is a 1 meter USB A B replacement cable for connection between your PC and the USB isolator...

Page 26: ...fer to your plug in module instructions for wiring Accuracy Better than 0 05 of span typical relative to the installed module plus an additional 0 05 of span relative to its operation on this carrier...

Page 27: ...12 20mA 20mA 59577 5 0V 10 0V 62556 5 0V 10V 65535 5 5V 11 0V 5 5V 11V 24mA 24mA RES 1 59577 1 59577 1 59577 1 59577 1 54612 1 43689 1 lsb 83 925uV 167 8uV 167 8uV 335 7uV 0 34132uA 0 34132uA Span 0 0...

Page 28: ...sourced from module Channel BLE Blue Channel Blue ON indicates Linking to channel is enabled for uBxx B Bluetooth wireless technology enabled models via the BLE DIP switch of the channel This LED is...

Page 29: ...o an Android smartphone tablet The USB interface is used to configure and calibrate the input of the module plugged into this carrier and each output of this splitter Because the input module power an...

Page 30: ...equivalent 1 59W Typical MAX TOTAL w uB42 or equivalent 1 96W Typical 1 Note Incremental power consumption of current output is 0 02W mA POWER REQUIRED BY FEATURE BY MODULE TYP MODEL BASE V OUT I OUT...

Page 31: ...ses Operating Temperature 40 C to 75 C 40 F to 158 F Storage Temperature 40 C to 85 C 40 F to 185 F Relative Humidity 5 to 95 non condensing Isolation Input field isolation is provided by the plug in...

Page 32: ...than pollution degree 2 as defined in EN IEC 60664 1 2 The equipment shall be installed in an enclosure that provides a degree of protection not less than IP 54 and only accessible with the use of a t...

Reviews: