Input Connections...
TB1 at the bottom edge (input side) of this carrier carries the field INPUT ± (TB1 lower
row), and EXC ± (TB1 back row) connections. Refer to your module instructions to make
wired connections and observe proper polarity. Be sure to ground your input as shown in
your
module’s respective connection
drawing (refer to your module manual instructions
listed below).
REFERENCE
DESCRIPTION
8501-037
uB31 & uB41 Voltage Input User Manual
8501-040
uB30 & uB40 mV Input User Manual
8501-041
uB32 DC Current Input User Manual
8501-042
uB42 DC Current Input w/2-Wire Excitation User Manual
8501-043
uB34 2/3-Wire & uB35 4-Wire RTD Input User's Manual
8501-044
Introduction to Agility Configuration Tool for microBlox®
8501-047
uB37 & uB47 TC Input User Manual
8501-050
uB04/uB08/uB16 Back-Panel User Manual
Power Connections
Power input to the carrier is safety isolated (250VAC) and reverse polarity protected up to
60V. Connect DC power supply from 6-32V at TB3 as shown below while observing
proper polarity, using 14 AWG wire rated for at least 85°C. Do not exceed 36V DC peak.
IMPORTANT:
It is recommended that supplies capable of delivering more than 2.5A of
power to the carrier be fused with a surge tolerant fuse. Unit includes a transient voltage
suppressor clamp across its power input that will clamp overvoltage conditions. A
sustained over-voltage condition from an unfused power supply could drive excessive
fault current flow that can damage the carrier if allowed to exceed 3A.