Model DT33x-0700
4-Wire DC I/V Dual Transmitter w/USB
Acromag, Inc. Tel: 248-295-0880
- 19 -
http://www.acromag.com
- 19 -
https://www.acromag.com
Configuration
Note that you should already have power connected to the
transmitter at this point, as this model does not draw power from
USB and you will not be able to configure, calibrate, or test the unit
without power applied.
After executing the Acromag Configuration software for this model,
the screen shown at left will appear, if you have not already
connected to your transmitter via USB (note red status message and
fields are blank under these conditions).
Connect your PC to the unit via USB, and the unit
’s model
-serial
information will appear in the Device field as shown in the second
screen at left.
If you happen to be connected to more than one unit via a USB hub,
you can use the Device scroll field to the right of the Device field to
select another unit (same model) using the serial information suffix
of the Device model data to discern one unit from another.
Once you have selected a device, click the
[Open]
button to open
communication with the unit.
With USB connected, after clicking [Open] of the Connection screen,
the selected unit’s Model, Serial Number, Manufacturer
will be
displayed as shown in the screen at left. Additionally, the
communication
Status field indicates “Device opened successfully”
as shown.
TIP:
Always close a connection with one device before selecting
another device and make sure that you have the correct model
software for the device(s) you are connected to.
At this point, you can click the
“Config
uration
”
tab to begin
configuring the unit,
“Calibration”
to calibrate an input or output, or
“Diagnostics”
to optionally test its operation.