Quick Overview
–
Windows
Click
“Open”
to connect to the DT336-0700 and your
screen will look like:
HELP
–
You can press F1 for Help on a selected or highlighted
field or control. You can also click the [?] button in the upper-
right hand corner of the screen and then click again to point
to a field or control to get a Help message pertaining to the
item you pointed to.
For detailed configuration and calibration procedures, see
the Operation Step-By-Step section of the Technical
Reference on page 18 of this manual.
This dual transmitter can be
configured and calibrated via its
USB Configuration Software and
a USB connection to your
Windows PC or laptop. The USB
software can be downloaded
free of charge from our web site
and is included on a CDROM
bundled with the Configuration Kit TTC-SIP (see
Accessories section). For this model, look for the
program DT33XConfig.exe. This software is compatible
with v7 or later versions of the Windows OS.
The initial USB configuration software screen for this
model is shown at left. Configuration information is
divided across four separately tabbed pages as follows:
Communication Set up, I/O Config/Test, Calibration, and
Diagnostics. A short description of each of these
configuration pages follows:
Communication Set up (First Connect to Unit Here)
•
Select from connected transmitters and Open/Close
communication with them.
•
Display the Model, Serial Number, and Manufacturer
of the connected transmitter and report the status
of communication with it.
Configuration (Configure the Unit Here)
•
Click
[Get I/O Config]
to retrieve the I/O
configuration of the current connected unit.
•
Select the operating Mode: Dual transmitter, Single
transmitter, or Signal Splitter.
•
Select the Input Range for each channel at TB1 and
TB3. You can select current ranges ±20mA, 0-20mA,
4-20mA, 0-11.17mA, or ±1mA or voltage ranges
±0.5V and 0-500mV for the DT336-0700 model.
•
Set the level of digital filtering to High, Medium,
Low, or None (No digital filter). I/O response times
vary with filter and are indicated next to your
selection (see Specifications for details).
•
Set the current or voltage Output Range to ±10V,
±5V, 0-5V, 0-10V, ±20mA, 0-20mA, or 4-20mA.
•
View the config message status in the Status field.
•
Use the I/O Scaling fields to specify the input range
endpoints to correspond to the nominal output
range zero and full-scale endpoints (some
over/under-range is included).
•
After making changes, send your settings to the unit
by clicking
[Send I/O Config]
and follow the on-
screen prompts.