7.2 The wireless communication of the instrument is faulty.
Troubleshooting suggestion: Please use the USB to 485 serial cable to connect to the RS485 interface of the
instrument first, read the parameters in the meter through communication, and confirm whether the parameters
in the meter are the same as the wireless configuration of the upper master station (channel and spreading
factor), if different, please modify The wireless parameters of the meter are consistent with the master station
and then re-test; if they are the same, it may be that the meter and the master station are too far away or the
on-site interference is serious. At this time, you can try to use an external suction cup antenna, or consider
adding a nearby wireless master station. Test again.