The primary application for AccuNet is in laboratories that have multiple VAV fume
hoods which require the total lab exhaust airflow to be instantly tracked by the supply
airflow to maintain the proper room pressurization. AccuValves must be equipped with
the optional AccuNet Module to operate on the AccuNet Network. Reference Appendix
C for detailed field instructions for configuring the AccuNet Network.
HELP
About
Close Active Window After Update
➢
When checked, this will close the active window after an update is made.
Display Comm Data
➢
When checked, this will display the communication data on the dashboard
task bar.
6.5.7 Dashboard Task Bar
Provides the USB communication status, device serial number, device tag and BACnet MAC
address at the top of the Dashboard.
6.5.8
AccuNet
AccuNet™ is a high
-speed serial bus that is designed to provide a room-level network for
summing the airflow volume of multiple AccuValves while providing a single analog signal
representing the total sum of the AccuValve airflows.
The primary application for AccuNet is in laboratories that have multiple VAV fume hoods
which require the total lab exhaust airflow to be instantly tracked by the supply airflow to
maintain the proper room pressurization. AccuValves must be equipped with the optional
AccuNet Module to operate on the AccuNet Network. Reference Appendix C for detailed field
instructions for configuring the AccuNet Network.