2. Error & Repair
Error
Cause
Repair
In case that the
P.B.F. is too low or
too high
• Measuring method.
Measuring is done under
unstable hydration state
such
as
just
after
physical
exercise,
a
bath, sweating, drinking
or overeating.
• Move or speak during
measurement
• Electrode holders or
measuring
parts
are
dirty
• Faulty in impedance
measuring
• Measure again in stable physique.
• If there is movement or speaking during
measurement, the weight and impedance are
not correct. Measure again correctly.
• Try to measure again after cleaning the
electrode holders with soft gauze.
• Try again after cleaning the hands and soles.
• Check if foreign materials are between
electrodes and measuring parts.
• Try again in correct posture and hold the
electrodes according to the measuring method.
Machine
doesn’t
Work When ‘start’
button is pressed
• When the electrode
holders
are
disconnected from the
main body or contact is
faulty.
•
'Start'
button
is
defective.
• make sure that the electrode holders are
connected to the main body well.
• If the same error occurs even after connecting
the electrodes well, please contact with company
or its local distributor from where device is
purchased.