58
ERROR & REPAIR
1. Kinds of Error & Repair
Error
Cause
Repair
Out of
range of
impedance
When the subject’s body impedance
deviates from the limit
- Insufficient touch to electrodes
- Impedance is out of range
- Range: 100 ~ 950 Ω
• Clean the measuring parts (the electrodes,
palms, and soles) and try again.
• Measure again with correct posture.
• Do not move during measurement.
• If the same error is repeated, please contact
our company or its local distributor from
where this device is purchased.
Out of
range of
body fat
When the subject’s P.B.F. deviates from
the limit
- Incorrect input of personal data
- P.B.F. is out of range
• Clean the electrode holders and try again.
• After checking that there is neither
something with wrong input of personal data
(age, gender) nor with measuring error of
weight and impedance, try again.
• It can't measure if the P.B.F. is out of range.
• When the same error occurs even after re-
measurement, please contact our company
or its local distributor from where the device
was purchased.
Out of
range of
measureme
nt
When the subject’s fatness is deviated
from the limit
- Mechanical error
• Input height correctly or if installed height
already, measure again.
• Confirm to measure weight and try again
correctly.
• It can't measure if the fatness is out of range.
• When the same error is occurred even re-
measurement, please contact with our
company or its local distributor where is
purchased.
Insufficient
electrode
connection
to Ankle
• When even one of all 4 electrodes
connects with ankle insufficiently
• Connect to ankle all 4 electrodes once more.