13
6. Guidance for Electromagnetic compatibility (EMC)
Details about the electromagnetic compatibility (EMC) of the ACCUNIQ BC360 are given below.
Before using the ACCUNIQ BC360, be sure to read and understand the following information.
1) Guidance and manufacturer’s declaration – electromagnetic emissions
The ACCUNIQ BC360 is intended for use in the electromagnetic environment specified below. The
customer or the user of the ACCUNIQ BC360 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11
Group 1
The ACCUNIQ BC360 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
The ACCUNIQ BC360 is suitable for use in all
establishments,
including
domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.