26
5. Troubleshooting and Maintenance (cont.)
5.4 IR Sensor Calibration (cont.)
5. Close the compartment, leaving the card inside the detection channel.
6. Press Menu to enter the set-up menu and use the navigation keys to
scroll and select the sensor adjustment mode. (SENSOR ADJUST)
7. Use the navigation keys to select the “STATIC SENSOR CALIBRATION”
mode and press “OK”.
8. Within the sensor calibration function (SENSOR CALIBRATION) press
“OK” to start the calibration. The equipment will confirm after a few
seconds if the calibration was successful or not.
9. If the calibration was not successful, repeat the calibration by
pressing “OK” one more time. If the error persists, check if the
calibration card is properly placed over the sensors inside the
equipment.
10. Turn off the equipment, open and remove the calibration card after
you have finished calibrating the sensors.
*If the calibration continues to be unsuccessful or the issue persists,
please contact our Technical Support team for assistance.