EXI-600 INVERTED MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
23
6.
OPERATION
6.1.
COMPONENT ADJUSTMENTS
6.1.1.
Trinocular observation tube adjustments
Eyepiece diopter adjustment
Set the diopter adjustment on each
eyepiece by rotating the eyepiece
while holding the barrel. Align the
“0” mark with the white dot.
Focus on a specimen on the stage
until the image in one eyepiece is
perfect.
Adjust the diopter of the other
eyepiece until the image through
that second eyepiece is good.
The eyepieces can be adjusted by
a maximum of ±5 diopters.
Adjust interpupillary distance
While viewing a sample, grasp and
adjust the eyetubes up or down at
the binocular hinge to increase or
decrease the interpupillary
distance (distance between the
pupils of the observer’s eyes).
Proper interpupillary distance
results in a single uniform view of
the specimen.
Interpupillary distance range:
55~75mm
Eyepiece
Barrel
Indicator
Interpupillary
distance
scale