5.2.10.
Fluorescence filter cube turret (optional)
Install the filter cubes into the cube
turret.
Insert the bottom of the “wings” of
a cube into the metal clip.
Gently push onto the back of the
cube until the top of the wings click
into the upper metal clips.
NOTE: When installed correctly,
there is no filter in the upper face
of the filter cube (filters occupy the
front and bottom positions, only, in
addition to the beamsplitter).
To insert the cube turret into the
frame, turn the locking lever down.
Raise the cover to reveal the
location into which the turret will be
inserted.