CLEANING
 

PART 

CLEANING 
AGENTS 

TIPS AND PRECAUTIONS 

Interior 
and Door 
Liners 

Soap and water 
Baking soda and 
water 

Use 2 tablespoons of baking 
soda in 1 quart of warm water 
Be sure to wring excess water 
out of sponge or cloth before 
cleaning around controls, light 
bulb or any electrical parts. 

Door 
Gaskets 

Soap and water 
 

Wipe gaskets and their seating 
surfaces with a clean soft cloth 

Shelves 

Soap and water 

Do not wash removable shelves 
in dishwasher 

Exterior 
and 
Handles 

Soap and water 
Non Abrasive 
Glass Cleaner 

Do not use commercial 
household cleaners, ammonia, 
or alcohol to clean handles 
Use a soft cloth to clean 
smooth handles 
Do not use a dry cloth to clean 
smooth handles 

 
Clean the glass with a mild detergent and water on a soft cloth or sponge. Rinse 
with water and wipe dry.  

 

For Sliding door units, dust and debris can begin to build up in the door track. 
To clean the door track  the doors should be removed and the track cleaned 
with mild detergent and water  on a soft  cloth or sponge.  Rinse  with  clean 
water and wipe dry. Contact technical support if you have difficulty removing 
the doors.  
 

Summary of Contents for ABT-FRP-04

Page 1: ...or Temp Control V5 2 Table of Contents RECEIVING AND SHIPPING DAMAGE HANDLING 2 SAFETY 3 RELEASE OF LIABILITY 3 ELECTRICAL INFORMATION 4 INSTALLATION 5 OPERATION 7 Quick Troubleshooting Guide 16 Maint...

Page 2: ...eir contact information can be found on the carrier s web site The second type of shipping damage is concealed damage This type of damage will probably not be apparent at time of delivery and may not...

Page 3: ...reezer This is a flammable material storage refrigerator or freezer All laboratory safety rules must be followed during the operation of this refrigerator or freezer RELEASE OF LIABILITY Before you st...

Page 4: ...to NEC National Electrical Code Consult the cabinet Serial Data plate for voltage cycle phase and amp requirements before making connection SUPPLY VOLTAGE SHOULD NOT VARY MORE THAN 5 FROM SERIAL PLAT...

Page 5: ...ntific medical application Many components are heavy duty and extra sized in order to meet the ultimate temperature performance Therefore the sounds generated from its operation may not be accepted by...

Page 6: ...IELD INSTALLATION Each refrigerator or freezer is equipped with a 3 8 probe access port hole for your independent probe installation The port hole is generally located in the back of your refrigerator...

Page 7: ...p alarm function D 100 temp display and alarm system is specifically designed to display the refrigerator s interior sample temperature and give out alarms if an error occurs FEATURES Sample temperatu...

Page 8: ...tteries will support the alarm system and the temperature display for 8 hours When the power resumes the batteries will be automatically recharged During a power failure event the visual display will...

Page 9: ...point MENU Press MENU button to enter the user adjustable settings Press UP or DOWN to select which setting you would like to make an adjustment CODE DESCRIPTION FACTORY DEFAULT ALU High temp alarm p...

Page 10: ...unction allows users to adjust the time delay between the occurrence of an error and the visual audible alarm s activation For example if the setting is 1 minute when the temp warms above the high tem...

Page 11: ...ome time to correct an error before sending the alarm to the remote monitor system For example rD is set to be 20 minutes A user forgets to close the refrigerator s door and the temperature warms up t...

Page 12: ...d allows the user to easily select the display units set point and differential set point Please Note The digital temperature controller has been factory set and tested to allow your unit to operate a...

Page 13: ...urn on and off in order to maintain the cold temperature in the storage chamber In this controller the point where the compressor is cut off is called SET POINT The point where the compressor is turne...

Page 14: ...t that SET or ALL could reach 1 C or 34 F refrigerator 30 C or 22 F freezer Change the set point compressor turn off point Press and hold SET until C or F icon blinking Press UP or DOWN to change the...

Page 15: ...e until Hy appears on the display Press UP or DOWN until Lod shows up Press SET to enter the setting Press UP or DOWN to toggle between the air temp P1 or the sample temp P3 Press SET once to confirm...

Page 16: ...refrigerator or freezer may not be able to maintain the interior temp at proper range Door is not closed properly Amount of stored product is overloaded Product replacements are pushed against rear wa...

Page 17: ...ther increases condensation Make sure there is a water trap U shaped loop in the drain tube near the compressor This will trap a small amount of water in the loop and prevent air from entering the cha...

Page 18: ...3 Make sure there is a water trap U shaped loop in the drain tube near the end This will trap a small amount of water in the loop and prevent air from entering the chamber through the tube BEFORE CALL...

Page 19: ...helves in dishwasher Exterior and Handles Soap and water Non Abrasive Glass Cleaner Do not use commercial household cleaners ammonia or alcohol to clean handles Use a soft cloth to clean smooth handle...

Page 20: ...ondenser clean you will minimize your service expense and lower your electrical costs The condenser requires scheduled cleaning every 30 day or more frequently in dirty environments Proper cleaning in...

Page 21: ...re to replace the louvered grill as it provides protection for the condenser 8 Reconnect the electrical power to the unit Hydrocarbon Service Notes According to U S Code of Federal Regulation 40 Part...

Page 22: ...r repaired parts will be warranted for only the unexpired portion of the original warranty Horizon Scientific Inc will not assume any shipping or cartage costs for parts under warranty These costs sha...

Page 23: ...conformities from any cause Purchaser further acknowledges that the purchase price of the Product reflects these warranty terms and remedies ALTERATION NEGLECT ABUSE MISUSE ACCIDENT DAMAGE DURING TRA...

Page 24: ...er at 800 648 4041 Horizon Scientific Inc s shipping records showing date of shipment shall be conclusive in establishing the warranty period All claims should include model number of the refrigerator...