Q:
“What if the instrument indicates low flow when the input
pressure is adequate?”
A:
#1 The needle may be clogged.
Use the following procedure to clean the needle
assembly and sensor chamber. For assistance, call AZI Customer Service at 800-528-7411 or
602-470-1414.
•
Set instrument oven temperature to 80 °C or below.
•
Obtain several ounces of isopropyl alcohol (IPA).
•
Fill a sample bottle ½ to ¾ full with IPA; fit with a cap and septum.
•
Place an empty sample bottle at the outlet port at the back of the instrument to
catch expelled solvent.
•
From the MAIN TEST SCREEN, navigate MENU
DIAGNOSTIC
MENU
MANUAL FLOW CONTROL MENU, and switch the MANUAL
FLOW CONTROL to BOTTLE PURGE.
•
After 5-10 seconds, solvent should begin to be expelled from the outlet port at the
back of the instrument.
•
When solvent expulsion slows or stops, switch MANUAL FLOW CONTROL to
TRANSPORT OUT.
•
Remove the sample bottle with the remaining solvent from the sample transport.
•
Switch MANUAL FLOW CONTROL to MANUAL OFF.
•
Escape to the MAIN TEST SCREEN by press ESC three times.
•
Wait for the Status Bar to read “READY TO TEST” before resuming normal
operation. This typically takes 15 to 30 minutes.
A:
#2 There may be a leak in the flow system.
If cleaning the needle according to the
above procedure did not resolve the issue, there may be a leak in the flow system. Call AZI
Customer Service at 800-528-7411 or 602-470-1414 for assistance.