Page 20
Turn the
VAPOR PRO
®
Rx
ON. From the Main Test Screen, press
[Menu]
to bring up the
MAIN MENU. Press
[Down]
(if necessary) so that CALIBRATION MENU is highlighted -
press
[Selct]
. At the Calibration Menu, press
[Down]
until TEMPERATURE CALIBRATION
is highlighted - press
[Selct]
.
Highlight ENTER DEVICE DATA and enter the appropriate
data for the interface serial number, probe serial number and
calibration/certification date.
•
The interface and probe serial numbers and the
certification date are located on the temperature
calibration interface label (the black box).
Return to the TEMPERATURE CALIBRATION MENU;
highlight VERIFICATION SETUP and press
[Selct]
.
•
The purpose of the verification function is to confirm
whether the calibration was successful or not at a
given set of temperatures. The verification process
can be run automatically following the calibration by
selecting ON at the AUTO-VERIFY prompt.
•
The rest of these instructions assume that the
verification is being run automatically. (AUTO-VERIFY: ON)
If you do not wish to
edit the temperatures, press
[Esc]
and move to the next step.
To select which temperature(s) will be verified, press
[Down]
to the TEMP # prompt and press
[Edit]
.
•
Using the
[Incr]
and
[Decr]
keys, enter the temperature you wish to verify and press
[Quit],
then
[Accpt]
. The temperatures selected should be the ones that best
represent those used for your product(s). When finished, press
[Esc]
.
Place a bottle with cap and clean septa on the transport.
Highlight RUN TEMP CAL and press
[Selct]
. The next screen will ask to ensure that you have
the temperature calibration device. If the device is hooked up, press
[Go]
.
When the verification is complete, the screen will display:
•
Verify: Passed or Verify: Failed
•
Key Options of
[Quit]
and
[Go]
Press
[Go]
to save the data.
If the unit failed at any given point, run the calibration again to improve the results.
If a printer is connected to the instrument, highlight PRINT TEMP CAL REPORT and press
[Selct]
to print out the calibration report.
Calibration is complete. Disconnect the calibration interface.