System Failure Codes
Occurring less frequently than condition prompts, system failure codes require the instrument to
be switched off and then powered back up. It is possible for a failure to occur due to a transient
condition. However, if the failure persists or occurs frequently, call Arizona Instrument
Customer Service at (800) 528-7411 or (602) 470-1414.
ERROR
ERROR MESSAGE TEXT
MOST LIKELY CAUSE
2
PC KEYBOARD COM FAIL
BAD CABLE OR CONFIG
3
PC KEY CONTROLLER TIMEOUT
DEAD KEYBOARD
5
RH SENSOR RANGE ERROR
FAILED RH SENSOR
6,7
FLOW SYSTEM ERROR
CABLE OR SENSOR FAIL
8
RH RTD RANGE ERROR
CABLE OR RTD FAIL
9
RH RTD RATE ERROR
Sensor Block HEATER
10
RH RTD TRACK ERROR
Sensor Block RTD OUT OF POSITION
11
BOTTLE RTD RANGE ERROR
CABLE OR RTD FAIL
12
BOTTLE RTD RATE ERROR
BOTTLE HEATER FAILURE
13
BOTTLE HEATER OVERTEMP
BOTTLE HTR RTD OUT OF POSITION
14
BOTTLE HTR CONTROL ERR
ROOM TOO COLD
15
Sensor Block RTD RANGE ERROR
CABLE OR RTD FAILURE
16
Sensor Block RTD RATE ERROR
Sensor Block HEATER FAILURE
17
Sensor Block HEATER OVERTEMP
Sensor Block RTD OUT OF POSITION
18
Sensor Block HTR CONTROL ERR
Sensor Block HTR OUT OF POSITION
19
A2D INIT FAILED
COMPONENT FAILURE, MAIN BOARD
20
A2D TIMEOUT ERROR
COMPONENT FAILURE, MAIN BOARD
21
MCU CLOCK FAIL ERROR
COMPONENT FAILURE, MAIN BOARD
22
COP TIMEOUT ERROR
COMPONENT FAILURE, MAIN BOARD
23
ILLEGAL OP CODE TRAP
COMPONENT FAILURE, MAIN BOARD
24
EPROM CHECK SUM FAIL
COMPONENT FAILURE, MAIN BOARD
25
SRAM CHECK FAIL
COMPONENT FAILURE, MAIN BOARD