7.
DRYNESS TEST
For all tests using a bottle, ENSURE the slick Teflon
®
side of the septum is to the inside of the
bottle or the moisture trapped in the septum will cause false moisture indications.
When the display on the Main Test Screen indicates
“Ready to Test”:
•
Press
[Memry]
•
Then select
DRYNESS TEST
Press
[START].
•
This test will run for two (2) minutes to test
for moisture that may have migrated into the
system.
•
If this is the first test after changing a portion
of the flow system; tubing, needle, filter, etc.,
or if the result is more than 5
µ
g, repeat the
Dryness Test up to four (4) times.
•
If the result does not fall under 5
µ
g after the
4th test:
Check for moisture in the dry air supply line.
Check and/or change the hydrophobic filter, see page 44.
Call AZI Customer Service at (800) 528-7411 or (602) 470-1414.