A.3
Impella
®
RP with the Automated Impella
®
Controller
MENU
The
MENU
soft button opens a menu of options related to controller settings, alarm history,
repositioning, offset adjustment, and starting a procedure. The menu includes the following
options:
•
Settings / Service
Service
System Information.
Opens the System Information table. This provides
information about the software version, IP addresses, current type of Impella
Catheter, and current catheter runtime.
Set Date/Time.
Displays the menu for changing the date and time
Service Timers.
Displays the Service Timers menu. Console operating time and
purge motor operating time are displayed in hours.
Screen Brightness.
Opens the Screen Brightness selection box. The brightness
of the screen display can be set from 50% to 100%. Select
OK
to confirm selection.
Select
Cancel
to cancel selection.
Language.
Opens the Language selection box. Use the selector knob to select
German, English, French, Italian, Spanish, or Dutch. The system will immediately
change the language on the controller for all displayed text. This language will be
used after system restart unless another language is selected.
Log Export.
Displays the Log Export menu for exporting all logs to a USB stick.
Disable (Enable) Placement Monitoring.
Disable (Enable) Reverse Flow Control.
Disable (Enable) Automatic Zero.
Disable (Enable) P-Level Suction Control.
Disable (Enable) Audio – Impella Sensor Failure.
Allows you to enable or
disable audio for the Impella Sensor Failure alarm. This selection is available only if an
Impella Sensor Failure alarm is active or the audio has been disabled for this alarm.
Disable (Enable) Audio – Purge Flow Low/System Blocked.
Allows you to
enable or disable audio for the Purge Flow Low or Purge System Blocked alarms. This
selection is only available if one of these two alarms is active or the audio has been
disabled for one of these alarms.
•
Alarm History
—opens the Alarm History table. This provides a visual display of the
chronology of stored alarm messages. The most recently occurring alarm message is
displayed at the top of the list. For each message, the date and time it occurred and the
alarm message heading is displayed. You can use the selector knob to select individual
alarm messages and an explanation for the selected alarm message will be displayed in
the failure description box. Press
EXIT
to exit the alarm history analysis.
•
Start Timed Data Recording
—starts the timed data recording function to save
real-time operating data for later analysis.
•
Start Manual Zero
—opens the procedure for manually zeroing the differential
pressure sensor.
•
Case Start
—begins the case procedure. Case Start is described in section 5 of this
manual under “Case Start.”