![background image](http://html1.mh-extra.com/html/abbott/neurosphere-3875/neurosphere-3875_user-manual_3045650094.webp)
Table 12. Guidance and Manufacturer’s Declaration –
Proximity Fields
Proximity
Test
IEC 60601
Test Level
Compliance
Level
Electromag
-
netic
Environmen
t
Guidance
5240 MHz,
5500 MHz,
5785 MHz:
9 V/m @ 217 Hz
pulse
modulation
9 V/m
WARNING: Do not use portable RF communications
equipment (including peripherals such as antenna
cables and external antennas) closer than 30 cm
(12 inches) to any part of the device, including cables
specified by Abbott Medical. Otherwise,
performance degradation may occur.
Radio Frequency Information
The effective radiated power is below the limits as
specified in:
▪
USA: FCC 47 CFR Part 15:2012
▪
▪
Canada: RSS-210 Issue 8
▪
90