![background image](http://html1.mh-extra.com/html/abbott/i-stat/i-stat_technical-bulletin_450987003.webp)
A. Customizing the Handheld to Enable the Wireless Functionality Using the Handheld Keypad
1. Press to turn on the handheld.
2. Press to change screen to
Administration Menu
.
3. Press (Customization).
4. Press (Change).
5. Enter password and press , or just press
if no password is required.
6. Press (Analyzer).
7. Press twice.
8. Press (Wireless).
9. Press to change setting to
Enabled
.
10. The “Wireless” listing should now appear as
Enabled
.
11. Press to turn handheld off and save the settings. (If setting
is not shown as
Enabled
, return to Step 9).
B. Customizing the Handheld to Enable Wireless Functionality Using CDS Version 5
1. Click on
Main
➔
Open Administration Function
➔
Customization
.
2. Type in your password and click OK. The default password is the word
istat
.
Note: Abbott Point of Care Inc. recommends changing the default password.
MENU
4
1
➔
1
ENT
3
ENT
3
15:26 18JAN10
Administration Menu
1 - Analyzer Status
2 - Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Customization
Default0
1 – View
2 – Change
2
2
Art: 726025-01A
Rev. Date 24-NOV-14
3
Main
System Tools Window Help
Instrument/Location
Operator
Database Maintenance
Inventory
Customization
User Administration
Open Monitor
Open Administration
Function
Open Data Viewer
Open Report
Close
Exit
Alt+X
X
X
X
X
Password
System Tools Window
X
A password is needed to access this workspace.
Password:
OK
Cancel
Change
Customization
Default0
1 – Analyzer
2 – ID Entry
3 – Patient Tests
4 – QC Tests
5 – Results
6 – Password
7 – Restore Factory
Settings
Analyzer
Customization
1 Wireless
Disabled
Change
Customization
Wireless
1 – Disabled
2 – Enabled