background image

•  

5 - Receive Setup

: Pressing          puts the handheld in setup mode, so that the handheld can connect to the 

i-STAT 1 Wireless Setup Utility.

After pressing         , the user is asked for a password.  After entering the appropriate password or pressing

ENT

, the user will see a 

Waiting for Setup

screen.  Once the test is complete, a 

Setup Successful 

or

Setup Failed

message (with an error code) will appear.

Enter Password

_ _ _ _ _

Waiting For

Setup

. . . .

1 – Cancel

Setup

Successful

1 – Continue

Note:  Pressing 

1-Continue

after

the Receive Setup screen
navigates the user back to
the Wireless Menu.

5

5

     

Note 5.1:

The 

Receive Setup

function should only be used by IT personnel using the instructions

found in the 

Configuring Wireless Settings in an i-STAT

®

1 Wireless Analyzer

Technical

Bulletin (Art: 726066).

i-STAT is a registered trademark of the Abbott Group of Companies in various jurisdictions.

Art:  726025-01A

12

Rev. Date: 24-NOV-14

Wireless Module

Error Code Description

FCC ID: YOPGS1500M

Wireless module is not configured

GS000

DHCP Process Time Out

GS001

Connection to Access Point has failed

GS002, GS010

Failed to make a TCP connection to Data Manager Port 6004

GS003

Unable to receive setup from APOC Wireless utility ad hoc

GS020, GS021

Wireless configuration storage failed to erase/ write to flash

GS022,GS023

Wireless configuration file corrupt or empty

GS024, GS025

Device customization failed to enable wireless module

GS030

Failed Ping request

TIMEOUT

ERROR CODE MESSAGES

Summary of Contents for i-STAT

Page 1: ...yzer Customizing the handheld to enable the Wireless functionality Transmitting data wirelessly to the Data Manager Using the Wireless Utility Menu to View the Wireless Settings and Test the Wireless Connections For i STAT 1 Wireless Specifications see the Technical Bulletin i STAT 1 Wireless Analyzer Specifications Art 728644 For instructions on configuring an i STAT 1 Wireless Analyzer see the T...

Page 2: ...d by the i STAT 1 Analyzer Model 300 Please refer to the i STAT 1 System Manual for full details on cartridge testing Items needed for the method of powering the handheld are the same as for the i STAT 1 Analyzer two 9 volt lithium batteries or a pair of rechargeable batteries purchased only from Abbott Point of Care However the user should expect an approximate 30 reduction in the number of cartr...

Page 3: ...ick OK The default password is the word istat Note Abbott Point of Care Inc recommends changing the default password MENU 4 1 1 ENT 3 ENT 3 15 26 18JAN10 Administration Menu 1 Analyzer Status 2 Data Review 3 QualityTests 4 Customization 5 Set Clock 6 Transmit Data 7 Utility Customization Default0 1 View 2 Change 2 2 Art 726025 01A Rev Date 24 NOV 14 3 Main System Tools Window Help Instrument Locat...

Page 4: ...ecific location to which this Handheld is assigned 5 Once the Preferences screen opens click on the Instrument tab Click the box to Enable wireless communication 6 Click OK and answer YES to the question about changing the preferences Enable Customization Location based customization profiles A_10 10 90 17 A_10 10 90 24 A_10 10 90 31 A_10 10 90 32 X Location Enable Use Update i STAT Philips Prefer...

Page 5: ...le click on the desktop shortcut or Internet Explorer Favorites for i STAT Customization 2 Make sure the Enable Customization box has a check mark in it Also make certain that the Enable Updates box is checked for the particular location to which this i STAT 1 Handheld is assigned 3 If the location to which this handheld is assigned has a checkmark under Uses Default double click on the alphanumer...

Page 6: ... its customized schedule it will attempt to download wirelessly all unsent results immediately before the power down 2 If the handheld is placed in a downloader or downloader recharger the handheld will first attempt to download wirelessly If it is successful the download process is complete If the wireless download is unsuccessful the handheld will attempt to download using the downloader or down...

Page 7: ...the transmission is successful Note 3 1 there are two additional parameters on this page which may aid the users in troubleshooting and assessing the strength of the wireless transmission signal Signal indicates the strength of signal in one of three terms Weak Medium or Strong Additionally an actual numeric measurement of the signal is displayed If users are obtaining failed transmissions when th...

Page 8: ...itional parameters on this page which may aid the users in troubleshooting and assessing the strength of the wireless transmission signal Signal indicates the strength of signal in one of three terms Weak Medium Strong Additionally an actual numeric measurement of the signal is displayed If users are obtaining failed transmissions when the signal strength is weak they should re attempt the transmi...

Page 9: ...sion is successful Note 3 1 When wireless is enabled and configured on an i STAT 1 Wireless Analyzer wireless communication will be attempted first If wireless communication fails wired communication will then be attempted Note 3 2 If there are unsent results remaining in the handheld at the completion of a transmission attempt a Communication Ended message will appear on the handheld indicating h...

Page 10: ...ta Manager for the i STAT 1 handhelds DNS Domain Name Server server that translates domain names to IP Addresses MAC Address MAC address of the wireless module contained in the handheld Firmware the Firmware revision of the wireless submodule 1 Initializing Wireless Module 1 Cancel Wireless Menu 1 View Setup 2 Ping Server 3 Test Server 4 Reset 5 Receive Setup Status State Associated Signal Weak 75...

Page 11: ...onnection Successful or Connection Failed with error code will appear on the display following the test 4 Reset Pressing resets the wireless module The user will receive either a Reset Successful or Reset Failed message with an error code on the display Reset powers off and powers back on the wireless module It does not reset the settings to factory default Resetting Wireless Module 1 Cancel Reset...

Page 12: ... by IT personnel using the instructions found in the Configuring Wireless Settings in an i STAT 1 Wireless Analyzer Technical Bulletin Art 726066 i STAT is a registered trademark of the Abbott Group of Companies in various jurisdictions Art 726025 01A 12 Rev Date 24 NOV 14 Wireless Module Error Code Description FCC ID YOPGS1500M Wireless module is not configured GS000 DHCP Process Time Out GS001 C...