9.   Fold the snap closure over the sample well until it clicks into place. 

Press on the tab. Do not press directly over the sample well.

Pt: 12345

16:10 25JanYY

i-STAT CHEM8+

 

Na 

mmol/L

         141

mmol/L

    4 . 2

Cl 

mmol/L

    1 0 4

iCa 

mmol/L

    1 . 2 2

T C O 2

m m o l / L

  2 6

 

      Page

Pt: 12345

16:10 25JanYY

i-STAT CHEM8+

 

Glu 

mg/dL

         90

BUN

mg/dL

   

1 7

Crea 

mg/dL

   

1 . 0

Hct 

%PCV

   

4 2

H b

g / d L

  1 4 . 3 

AnGap 

mmol/L

       16

      Page

If all results are not displayed on the first 
screen, press      to page through all 
screens.  

Fold closure

Insert cartridge

Review Results

•   View results on the handheld.  

Report results as per your  
facility policy.

Patient Testing Procedure - continued

10.   Push the cartridge into the handheld port until it clicks into place.   

Do not remove the cartridge from the handheld until the  
“CARTRIDGE LOCKED” message is gone. For Hct testing, the 
handheld should remain level until a result is obtained. A level surface 
includes running the handheld in the downloader/recharger.

•   See 

i-STAT 1 System 

Manual for Waived Tests

 

if the results are not 
displayed or if results are 
questioned. 

11.   After reviewing results, discard gloves, tube, transfer device and cartridge in 

biohazard waste container.