Control Testing Frequency

• Test one cartridge from each lot in each shipment upon receipt.
•  Test a single cartridge from the refrigerator monthly. Select cartridge in the following order: 

CHEM8+, 6+, Crea, EC4+, Glucose, E3+

Prepare Control

• 

 Control  Ampule

:  Must be at room temperature for at least  

 

 

          

30 minutes

 before using it.

•  

CHEM8+ Cartridges

:  use the TriControl Level 1 control.

• 

 6+ & Crea Cartridges

: use the i-STAT or TriControl Level 1  

 

 

 

     control.

• 

 EC4+, E3+, Glucose Cartridges

:  use the i-STAT or TriControl 

Level 3 control.

Test Control

Control Testing Procedure 

Shake ampule

Break ampule

MENU

1

3

1. 

 

 Press            to turn on handheld.

2.  Press            to change screen to Administration Menu.

3.  Press            for the Quality Tests menu.

4.  Press            for Control.

5.   As prompted by the handheld, scan or manually enter the 

Operator ID and Control Lot number.

6.   Scan the barcoded cartridge lot number.  

7.  Remove cartridge from pouch.

8.   Shake the ampule hard for 10 seconds.

9.   Break the ampule open. Use an ampule breaker or 

protect fingers by covering the ampule with gauze.

Continued