10.   Withdraw the control fluid slowly into a 1 or 3 cc plain 

syringe until the syringe is half filled.  Dispense 3 drops  
of control fluid onto a gauze pad.

Draw fluid

Fill cartridge

Fold closure

Control Testing Procedure - continued  

Review Results

•  Compare results on handheld to the Value Assignment 

Sheet. 

•  If any result is out of range, repeat the control testing with 

a fresh ampule and syringe.

•   If any result is still out of range, do not use the cartridge 

lot, and call Technical Support at 800-284-0702,  
Option 1.

11.   Fill the cartridge with the control fluid. Fold the  

closure over the sample well until it snaps into place.

12.   Push cartridge into the handheld port until it clicks into place, and do not remove until the 

“CARTRIDGE LOCKED” message is gone.   