Reviewing Test Results

n

    

The 0 key can be used to backlight the display to view
results in dim lighting.  (The backlight turns off after 90
seconds or when the 0 key is pressed again.)

n

    

Test results are displayed numerically and with bar graphs.
Tick marks indicate the reference ranges on the bar
graphs.  (Blood gases and their associated calculated
values are not displayed with bar graphs and reference
ranges.)

n

    

Test results are displayed for 2 minutes or a customized
time.  To recall the last set of results to the screen, turn the
handheld on and press 1 for Last Result.

n

    

To review results from the same patient, when results are
displayed, press 1 for Test Options and then 3 for History.
Scroll through test records using the 1 and 2 keys.

n

    

To review another patient’s results, turn the handheld on
and press the Menu key followed by the 2 key for Data
Review and the 1 key for patient.  Scan or enter the
Patient’s ID number.  Use the 1 and 2 keys to scroll
through the test records.  Or, press the Menu key followed
by the 7 key for List.  Select the test record(s) to be 
reviewed and press the Enter key.

Reportable and Reference Range

Measured:

Reportable

Reference Range

Test

Units

Range

(arterial)

(venous)

Sodium/Na

mmol/L (mEq/L)

100 – 180

138 – 146

138 – 146

Potassium/K

mmol/L (mEq/L)

2.0 – 9.0

3.5 – 4.9

3.5 – 4.9

Chloride/Cl

mmol/L (mEq/L)

65 – 140

98 – 109

98 – 109

Glucose/Glu

mmol/L

1.1 – 38.9

3.9 – 5.8

3.9 – 5.8

mg/dL

20 – 700

70 – 105

70 – 105

g/L

0.20 – 7.00

0.70 – 1.05

0.70 – 1.05

Lactate/Lac

mmol/L

0.30 – 20.00

0.36 – 1.25

0.90 – 1.70

mg/dL

2.7 – 180.2

3.2 – 11.3

8.1 – 15.3

Creatinine/Crea

mg/dL

0.2 – 20.0

0.6 – 1.3

0.6 – 1.3

µmol/L

18 – 1768

53 – 115

53 – 115

11

4. Press         for Simulator.

5.    Scan or enter Operator ID.

6.    Enter the Simulator ID (serial number).

7.    Insert the simulator into the cartridge port.

8.    View results on handheld’s screen.

9.    If PASS is displayed, continue to use the handheld.

10.  If FAIL is displayed for the external simulator, reinsert the simulator.

       If FAIL is displayed a second time, do not use the handheld and contact your

Support Services representative.

Thermal Probes and Room Temperature Checks

See System Manual for these quality assurance procedures that are performed once or
twice per year.

Verification of Cartridges Using 2 Levels of Specified Controls

Cartridge Type                                                                          Control Type

G, Crea, G3+, CG4+ 

TriControls 

or

i-STAT controls

CHEM8+, E3+, EC4+, 6+

TriControls

EC8+, EG6+, EG7+, CG8+

ACTk, ACTc

i-STAT ACT controls 

PT/INR

i-STAT PT/INR controls 

cTnI

i-STAT cTnI controls 

CK-MB

i-STAT CK-MB controls 

BNP

i-STAT BNP controls 

ß-hCG

i-STAT ß-hCG controls 

n

  

Verify that the storage conditions listed above have been maintained.

Cartridge

n

  

Immediately check temperature strip enclosed with each shipment of cartridges.
Follow the instructions on the card.

n

  

Verify the integrity of cartridges included in every shipment, upon receipt, by
analyzing 2 levels of appropriate controls (see table below) along with a representative
sample of each new lot and by comparing the results to the expected values
published in the Value Assignment Sheets.*  Any analyzer that has passed the
Electronic Simulator test may be used in the verification.

       * This information is not a manufacturer’s system instruction. It is a suggestion to

comply with regulatory requirements that may pertain to your laboratory.

4

Summary of Contents for i-STAT 1

Page 1: ...ch it is likely to recover Insert the Electronic Simulator If PASS continue to use handheld Analyzer Error See Manual Handheld detects problem from which it may not recover Insert Electronic Simulator...

Page 2: ...code repeats more than twice try another handheld Cartridge Preburst Calibrant pack burst before cartridge inserted into handheld Use another cartridge do not press on center of cartridge Check that c...

Page 3: ...of saline and the rst 5 mL of blood or six dead space volumes should be discarded c Withdraw the sample into a fresh plastic syringe with no anticoagulant and test immediately For extracorporeal line...

Page 4: ...interfere with cartridge tests 13 Cleaning the Handheld and Downloader Please see Section 17 of the i STAT 1 System Manual Routine Care of the Analyzer Downloader for instructions on cleaning and deco...

Page 5: ...pted 5 Scan or enter Operator ID 6 Enter the control lot number 7 Scan the lot number on the cartridge pouch 8 Fill a cartridge with the control and close the cover 9 Insert the cartridge into the car...

Page 6: ...2 11 3 8 1 15 3 Creatinine Crea mg dL 0 2 20 0 0 6 1 3 0 6 1 3 mol L 18 1768 53 115 53 115 11 4 Press for Simulator 5 Scan or enter Operator ID 6 Enter the Simulator ID serial number 7 Insert the simu...

Page 7: ...reinsert the cartridge or use an external simulator The external simulator check is performed as follows 1 Turn the handheld on 2 Press to access the Administration Menu 3 Press for Quality Tests MEN...

Page 8: ...Printer or Martel Printer Without Downloader or Downloader Recharger 1 Turn printer on if green power light is not on 2 Align IR windows of handheld and printer 3 Display results 4 Press 5 Do not move...

Reviews: