Date Invalid, 
Check Clock

Date outside six
month lifetime of
software.

Select 5-Clock Set from
Administration Menu. (Password
protected.)

Dead Batteries,
Replace Batteries

Insufficient power to
complete a test cycle.

Replace disposable batteries or
recharge the rechargeable battery.

Temperature Out of
Range, Check
Status page

Temperature outside
operating range of 16
to 30 °C.

Check handheld temperature by
pressing 1 for Analyzer Status under
the Administration Menu. Move
handheld to warmer area if below
operating range or to cooler area if
above the range.

Invalid or Expired
CLEW

Software expired or
corrupt.

Verify that the handheld’s date is
correct. Change software if expired.
Update again if not expired.

Analyzer Interrupted,
Use Another
Cartridge

Last cartridge run not
completed.

Check the battery pack inserted
properly. Check for Low Battery
startup warning.

Cartridge Error

Usually problem with
sample or cartridge
filling.

Use another cartridge. If same code
repeats more than twice, try another
handheld.

Cartridge Preburst

Calibrant pack burst
before cartridge inserted
into handheld.

Use another cartridge - do not press
on center of cartridge. Check that
cartridges have not been frozen.

Unable to Position
Sample

Cartridge not sealed.
Clot in sample.
Aberrant cartridge.

Use another cartridge.

Sample Positioned
Short of Fill Mark

Cartridge underfilled.

Use another cartridge - fill to Fill
Mark.

Sample Positioned
Beyond Fill Mark

Cartridge overfilled.

Use another cartridge - do not fill
beyond Fill Mark.

Test Cancelled by
Operator

User did not respond
to mandatory prompt
before handheld time
out.

No action required.

Cartridge Type Not
Recognized

Software does not
recognize cartridge.

Update software. Check to see if
cartridges are expired.

i-STAT 1

Downloader

Electronic

Simulator

i-STAT Printer

Martel Printer

See System Manual for instructions.

For

in vitro

diagnostic use.

2

Intended Use: The i-STAT 1 Analyzer is intended for use with i-STAT cartridges
for the 

in vitro 

quantification of various analytes in whole blood.

i-STAT 1

Downloader/Recharger

i-STAT Cartridges

System Components

15

Quality Check Messages and Codes

i-STAT 1 Analyzer

Summary of Contents for i-STAT 1

Page 1: ...ch it is likely to recover Insert the Electronic Simulator If PASS continue to use handheld Analyzer Error See Manual Handheld detects problem from which it may not recover Insert Electronic Simulator...

Page 2: ...code repeats more than twice try another handheld Cartridge Preburst Calibrant pack burst before cartridge inserted into handheld Use another cartridge do not press on center of cartridge Check that c...

Page 3: ...of saline and the rst 5 mL of blood or six dead space volumes should be discarded c Withdraw the sample into a fresh plastic syringe with no anticoagulant and test immediately For extracorporeal line...

Page 4: ...interfere with cartridge tests 13 Cleaning the Handheld and Downloader Please see Section 17 of the i STAT 1 System Manual Routine Care of the Analyzer Downloader for instructions on cleaning and deco...

Page 5: ...pted 5 Scan or enter Operator ID 6 Enter the control lot number 7 Scan the lot number on the cartridge pouch 8 Fill a cartridge with the control and close the cover 9 Insert the cartridge into the car...

Page 6: ...2 11 3 8 1 15 3 Creatinine Crea mg dL 0 2 20 0 0 6 1 3 0 6 1 3 mol L 18 1768 53 115 53 115 11 4 Press for Simulator 5 Scan or enter Operator ID 6 Enter the Simulator ID serial number 7 Insert the simu...

Page 7: ...reinsert the cartridge or use an external simulator The external simulator check is performed as follows 1 Turn the handheld on 2 Press to access the Administration Menu 3 Press for Quality Tests MEN...

Page 8: ...Printer or Martel Printer Without Downloader or Downloader Recharger 1 Turn printer on if green power light is not on 2 Align IR windows of handheld and printer 3 Display results 4 Press 5 Do not move...

Reviews: