Patient Test Procedures

Cartridge Test Procedure

• DO NOT open cartridge pouch before scanning the barcode.

1.  Press          to turn on handheld.

2.  Press          i-STAT Cartridge.

3.  Follow handheld prompts.

4.  Scan the lot number on the cartridge pouch.

Position barcode 3 - 9 inches from scanner 
window on the handheld.

Press and hold           to activate the scanner.

Align the red laser light so it covers entire barcode.

The handheld will beep when it reads the 
barcode successfully.

5.  Continue normal procedures for preparing the sample, filling and sealing cartridge.

2

SCAN

Laser Radiation – Do not stare into beam.  Class 2 laser product.
Laser Diode 650 nm Maximum Output 1.0 mW.

5

6.  Push the sealed cartridge into the handheld port until it clicks into place.  Wait for the test

to complete.

    

Note: For ACT, PT/INR, Hct and immunoassay testing, the handheld must remain on a level
surface with the display facing up during testing.

A level surface includes running the handheld in the Downloader/Recharger.

7.  Review results.

12

Procedure

1.    Turn the handheld on and press         to access the Administration Menu.

2.    Press         for Quality Tests.

3.    Press         for Control.

4.    Press         for i-STAT Cartridge (if prompted).

5.    Scan or enter Operator ID.

6.    Enter the control lot number.

7.    Scan the lot number on the cartridge pouch.

8.    Fill a cartridge with the control and close the cover.

9.    Insert the cartridge into the cartridge port.

10.  Enter chart page information if applicable.

11.  View results on handheld’s display.

12.  Remove and discard cartridge when Cartridge Locked message disappears.

13.  Press         for Test Options on the results page and press         for Next Level if

testing another level of control. 

3

1

1

1

1

MENU

Hardware Procedures

Replacing Batteries

1.    Slide the battery compartment door off.

2.    Tilt the handheld slightly to slide out the battery carrier.

3.    Remove the old batteries from the carrier and replace

with 2 new 9V lithium batteries.  

4.    Insert the carrier back into the compartment – label

facing up and electrical contacts first.

5.    Slide the battery compartment door into place.

Charging the Rechargeable Battery

Placing a handheld in a Downloader/Recharger will automatically initiate recharging of the
rechargeable battery.  The indicator light on top of the Downloader/Recharger will be
green (trickle charge), red (fast charge), or blinking red (fast charge pending) when a
handheld with a rechargeable battery is placed in the Downloader/Recharger.  

No damage will be caused if a handheld with disposable batteries installed is placed in
the Downloader/Recharger.

Summary of Contents for i-STAT 1

Page 1: ...ch it is likely to recover Insert the Electronic Simulator If PASS continue to use handheld Analyzer Error See Manual Handheld detects problem from which it may not recover Insert Electronic Simulator...

Page 2: ...code repeats more than twice try another handheld Cartridge Preburst Calibrant pack burst before cartridge inserted into handheld Use another cartridge do not press on center of cartridge Check that c...

Page 3: ...of saline and the rst 5 mL of blood or six dead space volumes should be discarded c Withdraw the sample into a fresh plastic syringe with no anticoagulant and test immediately For extracorporeal line...

Page 4: ...interfere with cartridge tests 13 Cleaning the Handheld and Downloader Please see Section 17 of the i STAT 1 System Manual Routine Care of the Analyzer Downloader for instructions on cleaning and deco...

Page 5: ...pted 5 Scan or enter Operator ID 6 Enter the control lot number 7 Scan the lot number on the cartridge pouch 8 Fill a cartridge with the control and close the cover 9 Insert the cartridge into the car...

Page 6: ...2 11 3 8 1 15 3 Creatinine Crea mg dL 0 2 20 0 0 6 1 3 0 6 1 3 mol L 18 1768 53 115 53 115 11 4 Press for Simulator 5 Scan or enter Operator ID 6 Enter the Simulator ID serial number 7 Insert the simu...

Page 7: ...reinsert the cartridge or use an external simulator The external simulator check is performed as follows 1 Turn the handheld on 2 Press to access the Administration Menu 3 Press for Quality Tests MEN...

Page 8: ...Printer or Martel Printer Without Downloader or Downloader Recharger 1 Turn printer on if green power light is not on 2 Align IR windows of handheld and printer 3 Display results 4 Press 5 Do not move...

Reviews: