CHEM8+ Cartridges

a.    CHEM 8+ cartridges require the use of whole blood collected in non-

heparinized evacuated tubes or syringes, as long as sample is tested
immediately upon draw, 

or;

b.    Heparinized whole blood collected in balanced heparin syringes or capillary

tubes, 

or;

c.    Heparinized whole blood collected in evacuated tubes containing lithium

heparin, as long as the tubes are filled to capacity.

n

    

Troponin I/cTnI, ß-hCG and CK-MB Tests

a.    cTnI and CKMB cartridges require the use of heparinized whole blood or

plasma samples collected in syringes or evacuated tubes containing lithium
or sodium heparin and filled to capacity, 

or;

b.    Non-heparinized whole blood or plasma samples tested within one minute

of drawing from a patient into a plastic syringe or plastic evacuated tube
containing no additives.

The use of whole blood or plasma samples containing other anticoagulants such as
EDTA, oxalate, and citrate will cause deactivation of the alkaline phosphatase,
resulting in decreased cTnI, ß-hCG or CK-MB readings.

Capillary tubes and direct skin punctures (e.g., fingersticks) should not be used
with the cTnI,  ß-hCG or CK-MB cartridge.

n

    

BNP Tests

BNP catridges require the use of EDTA whole blood or plasma samples collected
in plastic syringes or evacuated tubes containing EDTA and filled to capacity.

The use of whole blood or plasma samples containing other anticoagulants such
as oxalate and citrate is not recommended.

Capillary tubes and direct skin punctures (e.g., fingersticks) should not be used
with the BNP cartridge.

4

Limitations

Interfering substances in the patient’s sample may cause an increase or decrease in a
result.  Refer to the Cartridge and Test Information Sheets and Technical Bulletins for
substances and/or conditions that may interfere with cartridge tests.

13

Cleaning the Handheld and Downloader

Please see Section 17 of the i-STAT 1 System Manual (Routine Care of the Analyzer &
Downloader) for instructions on cleaning and decontaminating the handheld and
downloader.

Replacing Paper in the Martel Printer

1.    Squeeze the front and back of the paper cup to open.

2.    Remove remaining paper by pressing the Paper feed button.  Do not pull paper

through printer mechanism.

3.    Reel off a few centimeters of paper from the new paper roll and check that the

end has a clean straight edge.

4.    Slide the leading edge of the paper through the paper entry slot until you feel

resistance.  Paper feeds from underneath the roll.

5.    Press the Paper Feed button and feed the paper through the printer mechanism.

6.    Keep the Paper Feed button depressed until the paper passes through the

paper exit slot.

7.    Sit the new paper roll in the paper cup and close the lid.

Charging Rechargeable Battery in External Recharge Compartment

Placing a rechargeable battery into the recharging compartment will automatically initiate
trickle recharging.  The indicator light near the recharging compartment will be green when a
rechargeable battery is placed in the compartment.

STEP / ACTION

1     The battery pack has two labels:  one for orientation in the handheld and one for

orientation in the Downloader/Recharger. With the label with the Downloader facing
up and the electrical contact end of the pack facing the contacts in the battery
compartment, insert the pack into the compartment as shown on the label.

2     To remove the battery after it is charged, back the pack out of the compartment.

Full recharge from a discharged state takes approximately 40 hours.

Caution

If you are using rechargeable batteries, use 

only

rechargeable batteries and recharging

equipment supplied by your i-STAT distributor.  Other batteries and rechargers may affect test
results and pose other hazards to operators and patients.

Summary of Contents for i-STAT 1

Page 1: ...ch it is likely to recover Insert the Electronic Simulator If PASS continue to use handheld Analyzer Error See Manual Handheld detects problem from which it may not recover Insert Electronic Simulator...

Page 2: ...code repeats more than twice try another handheld Cartridge Preburst Calibrant pack burst before cartridge inserted into handheld Use another cartridge do not press on center of cartridge Check that c...

Page 3: ...of saline and the rst 5 mL of blood or six dead space volumes should be discarded c Withdraw the sample into a fresh plastic syringe with no anticoagulant and test immediately For extracorporeal line...

Page 4: ...interfere with cartridge tests 13 Cleaning the Handheld and Downloader Please see Section 17 of the i STAT 1 System Manual Routine Care of the Analyzer Downloader for instructions on cleaning and deco...

Page 5: ...pted 5 Scan or enter Operator ID 6 Enter the control lot number 7 Scan the lot number on the cartridge pouch 8 Fill a cartridge with the control and close the cover 9 Insert the cartridge into the car...

Page 6: ...2 11 3 8 1 15 3 Creatinine Crea mg dL 0 2 20 0 0 6 1 3 0 6 1 3 mol L 18 1768 53 115 53 115 11 4 Press for Simulator 5 Scan or enter Operator ID 6 Enter the Simulator ID serial number 7 Insert the simu...

Page 7: ...reinsert the cartridge or use an external simulator The external simulator check is performed as follows 1 Turn the handheld on 2 Press to access the Administration Menu 3 Press for Quality Tests MEN...

Page 8: ...Printer or Martel Printer Without Downloader or Downloader Recharger 1 Turn printer on if green power light is not on 2 Align IR windows of handheld and printer 3 Display results 4 Press 5 Do not move...