7

10

Transmitting Results

1.    Place handheld in Downloader or Downloader/Recharger.

2.    Do not move handheld until Communication in Progress message

disappears.

Storage Conditions and Preparation for Use

Cartridges

n

  

Store at temperatures between 2 and 8 °C (35-46 °F).  Do not use after expiration
date on cartridge pouch and box.

n

  

Equilibrate a single cartridge for 5 minutes or a box of cartridges for 1 hour at room
temperature before opening pouches.

n

  

Store cartridges at room temperature for the timeframe indicated on the cartridge box.
Mark the cartridge box or cartridge pouches with the room temperature expiration date.
Do not expose to temperatures above 30 °C (86 °F).  Do not return cartridges to the
refrigerator after room temperature equilibration.

n

  

Use cartridge immediately after opening pouch.  If the pouch has been punctured,
the cartridge should not be used.

Analyzer

n

  

Storage/Transport temperature: -10 to 46 °C (14-115 °F).

n

  

The handheld’s operating temperature range is 16 to 30 °C (61-86 °F).

n

   Store handhelds near the testing location or in an area close to the temperature of

the testing area.  Do not store handhelds near equipment that gives off heat or in
direct sunlight.

Quality Assurance

Handheld

Electronic Simulator

Perform an electronic check on each handheld in use once a day with either the internal or
external Electronic Simulator or as needed for regulatory compliance.  The internal simulator
check is initiated, every 24 hours or according to a customized schedule, when a cartridge is
inserted into the cartridge port. If the internal simulator result is PASS, the cartridge test
proceeds and the simulator results are stored.  If FAIL is displayed for the internal simulator,
reinsert the cartridge or use an external simulator.  The external simulator check is performed
as follows:

1.    Turn the handheld on.

2.    Press         to access the Administration Menu.

3.    Press         for Quality Tests.

MENU

3

Reportable and Reference Range 

(continued)

Measured 

(continued)

:

Reportable

Reference Range

Test

Units

Range

(arterial)

(venous)

pH

6.50 – 8.20

7.35 – 7.45

7.31 – 7.41

P

CO

2

mmHg

5 – 130

35 – 45

41 – 51

kPa

0.67 – 17.33

4.67 – 6.00

5.47 – 6.80

TCO

2

mmol/L (mEq/L)

5-50

23 – 27

24 – 29

(on CHEM8+ cartridge only)

P

O

2

mmHg

5 – 800

80 – 105

kPa

0.7 – 106.6

10.7 – 14.0

Ionized Calcium/iCa

mmol/L

0.25 – 2.50

1.12 – 1.32

1.12 – 1.32

mg/dL

1.0 – 10.0

4.5 – 5.3

4.5 – 5.3

Urea Nitrogen/BUN 

mg/dL

3 – 140

8 – 26

8 – 26

Urea

mmol/L

1 – 50

2.9 – 9.4

2.9 – 9.4

mg/dL

6 – 300

17 – 56

17 – 56

Hematocrit/Hct

%PCV

15 – 75

38 – 51

38 – 51

Fraction

0.15 – 0.75

0.38 – 0.51

0.38 – 0.51

Celite Activated

seconds

50 – 1000

74 – 125 (Prewarm)

74 – 125 (Prewarm)

Clotting Time/

84 – 139 (Nonwarm) 84 – 139 (Nonwarm)

Celite

ACT

The range from 80 - 1000 seconds has been verified through method comparison studies.

Kaolin Activated

seconds

50 – 1000

74 – 137 (Prewarm)

74 – 137 (Prewarm)

Clotting Time/

82 – 152 (Nonwarm) 82 – 152 (Nonwarm)

Kaolin

ACT

The range from 77 - 1000 seconds has been verified through method comparison studies

Prothrombin Time / PT

INR

0.9 – 8.0

Performance characteristics have not been established for INRs above 6.0

Troponin I / cTnI

ng/mL (µg/L)

0.00 – 50.0

0.00 – 0.03*

0.00 – 0.08**

Performance characteristics have not been established for cTnI values above 35.00 ng/mL.
* Represents the 0 to 97.5% range of results.
** Represents the 0 to 99% range of results.

Creatinine Kinase MB

ng/mL (µg/L)

0.0 – 150.0

0.0 – 3.5***

/ CK-MB

***Represents the 0 to 95% range of results.

ß-Type Natriuretic

pg/mL (ng/L)

15 – 5000

<15 – 50#

Peptide / BNP

# Represents the 0 to 95% range of results

Total Beta-Human

IU/L

5.0 – 2000.0

<5.0

Chorionic Gonadotropin
/ß-hCG

Celite is a registered trademark of Celite Corporation, Santa Barbara, CA. for its diatomaceous earth products.

Summary of Contents for i-STAT 1

Page 1: ...ch it is likely to recover Insert the Electronic Simulator If PASS continue to use handheld Analyzer Error See Manual Handheld detects problem from which it may not recover Insert Electronic Simulator...

Page 2: ...code repeats more than twice try another handheld Cartridge Preburst Calibrant pack burst before cartridge inserted into handheld Use another cartridge do not press on center of cartridge Check that c...

Page 3: ...of saline and the rst 5 mL of blood or six dead space volumes should be discarded c Withdraw the sample into a fresh plastic syringe with no anticoagulant and test immediately For extracorporeal line...

Page 4: ...interfere with cartridge tests 13 Cleaning the Handheld and Downloader Please see Section 17 of the i STAT 1 System Manual Routine Care of the Analyzer Downloader for instructions on cleaning and deco...

Page 5: ...pted 5 Scan or enter Operator ID 6 Enter the control lot number 7 Scan the lot number on the cartridge pouch 8 Fill a cartridge with the control and close the cover 9 Insert the cartridge into the car...

Page 6: ...2 11 3 8 1 15 3 Creatinine Crea mg dL 0 2 20 0 0 6 1 3 0 6 1 3 mol L 18 1768 53 115 53 115 11 4 Press for Simulator 5 Scan or enter Operator ID 6 Enter the Simulator ID serial number 7 Insert the simu...

Page 7: ...reinsert the cartridge or use an external simulator The external simulator check is performed as follows 1 Turn the handheld on 2 Press to access the Administration Menu 3 Press for Quality Tests MEN...

Page 8: ...Printer or Martel Printer Without Downloader or Downloader Recharger 1 Turn printer on if green power light is not on 2 Align IR windows of handheld and printer 3 Display results 4 Press 5 Do not move...