System Controller |
25
CAUTION:
The 11 V lithium-ion backup battery inside the
backup System Controller must be charged once
every 6 months.
Charge the backup battery
inside
the backup
System Controller.
1. Connect the backup System Controller to a
power source (MPU or two HeartMate 3™
LVAD 14 V lithium-ion batteries).
2. When the System Controller is connected
to power, the user interface screen displays
“Charging” or “Charging Complete.”
3. When charging is complete, perform a self test
on the backup System Controller (press and
hold the battery button for 5 seconds).
4. Disconnect power from the backup System
Controller. Place the backup System Controller
back in the protective case. No further action is
needed for 6 months.
SYSTEM CONTROLLER:
CHARGING THE
BACKUP BATTERY
IMPORTANT:
Do not remove power until “Charging
Complete” is displayed. It can take
up to 3 hours to charge the System
Controller’s backup battery.