24
| System Controller
SYSTEM CONTROLLER:
BACKUP BATTERY
An 11 V lithium-ion backup battery inside the
System Controller
provides at least 15 minutes
of backup power to the LVAD if the in-use power
source is disconnected or fails.
The System Controller’s backup battery will be
charged once external power is connected
to the System Controller (batteries or MPU).
WARNING:
The 11 V lithium-ion backup battery should be
used only for temporary support during a power-
loss emergency. Inappropriate use of this backup
battery may result in diminished run time during
a power-loss emergency.