System Controller |
17
HAZARD
SYSTEM
CONTROLLER
SCREEN
ACTIVE
SYMBOLS
ALARM MEANS
TO RESOLVE ALARM
Both power cables
are disconnected.
1. Immediately connect to a working
power source (MPU or two fully charged
HeartMate 3™ LVAD 14 V lithium-ion
batteries).
2. If alarm persists, call your hospital
contact immediately.
Low flow: Flow is
less than 2.5 liters
per minute.
Call your hospital contact immediately for
diagnosis and instructions.
Low voltage: Power
input is extremely
low, with less
than 5 minutes
remaining.
1. Immediately connect to a working power
source (MPU or two fully charged
HeartMate 3 LVAD 14 V lithium-ion batteries).
2. If alarm persists, call your hospital
contact immediately.
SYSTEM CONTROLLER:
ALARMS