10
| System Controller
1
SYSTEM CONTROLLER:
BUTTONS AND USER INTERFACE SCREEN
1. BATTERY BUTTON
• Displays the battery power
(press and release).
• Tests the controller (press and hold
for 5 seconds).
• Puts the controller into sleep mode
(press and hold for 5 seconds when
nothing is connected to the controller).
2. DISPLAY BUTTON
Activates the user interface screen
(press and release) to display information
about system operation.
3. SILENCE ALARM BUTTON
• Silences an active alarm (press and release).
• Displays previous alarms (press and
release the silence alarm and display
buttons simultaneously).
4. USER INTERFACE SCREEN
Displays information including pump speed,
pump flow, pulsatility index, power and
charge status of the backup battery.
3
2
4