![background image](http://html1.mh-extra.com/html/abbott/freestyle-libre-2/freestyle-libre-2_manual_3327752050.webp)
Reader Specifications
Blood glucose assay range
20 to 500 mg/dL
Blood ketone assay range
0 .0 to 8 .0 mmol/L
Reader size
95 mm x 60 mm x 16 mm
Reader weight
65 grams
Reader power source
One lithium-ion rechargeable battery
Reader battery life
4 days of typical use
Reader memory
90 days of typical use
Reader operating
temperature
10 °C to 45 °C
Reader storage
temperature
-20 °C to 60 °C
Operating and storage
relative humidity
10-90 %, non-condensing
Reader moisture
protection
Keep dry
Sensor transmission
range
6 metres (20 ft) unobstructed
Operating temperature
10 °C to 45 °C
Sensor Applicator and
Sensor Pack storage
temperature
4 °C to 25 °C
Operating and storage
relative humidity
10-90 %, non-condensing
Sensor water resistance
and ingress protection
IP27: Can withstand immersion into
one metre (3 ft) of water for up to
30 minutes . Protected against insertion
of objects >12 mm diameter
Operating and
storage altitude
-381 metres (-1,250 ft) to 3,048 metres
(10,000 ft)
Radio Frequency
2 .402-2 .480 GHz BLE; GFSK; 0dBm EIRP
91
92
ART41600-101_rev-A_manual.indd 91-92
ART41600-101_rev-A_manual.indd 91-92
10/7/20 2:55 PM
10/7/20 2:55 PM